Background: Boron Neutron Capture Therapy (BNCT) generally comprises of four dose components. The primary one which directly deals damage to the tumor is called the boron-dose, as it originates from the nuclear reaction of boron-10. While the, the other three came from the nuclear reaction of hydrogen and nitrogen atoms in the tissues and called the non-boron dose. These nuclear reaction produces secondary particles which stochastically behave when interacts with the tissue. As such, these occurrences pose a great challenge for determining/calculating accurate dose due to their stochastic nature and could deal damage to the healthy tissues. So, a Treatment Planning System (TPS) is utilized to help in determining an effective BNCT treatment. Such that, TPS helps to evaluate dose distributions and therapeutic effect which consequently helps to determine adequate treatment parameters to a successful BNCT treatment. However, TPS are not always compatible to any treatment and some TPS has particular conditions on when can it be utilizing into its full potential and when will it not be efficient for BNCT treatment planning/treatment. Therefore, this study briefly reviews some of the features, capabilities, and limitations of each major TPS’s namely, NCTPlan, SERA, JCDS, THORPlan, NeuCure, and Tsukuba Plan.
Conclusion: This study highlighted the strength and shortcomings of the major TPS’s in treatment planning for BNCT application. This paper gives insights on the future research to be undertaken of the development of TPS for the advancement of BNCT.