Odette Ngano Samba, Chi Emmanuel *, C. Tetchoka Manemo and L. C. Fai
Department of Physics, University of Dschang, Dschang, Cameroon
Received Date: 2022-10-26 | Accepted Date: 2022-10-25 | Published Date: 2022-10-25Odette Ngano Samba, Chi Emmanuel *, C. Tetchoka Manemo and L. C. Fai
Department of Physics, University of Dschang, Dschang, Cameroon
Received date: Septmber 26,2022, Manuscript No. IPIMP-22-14074; Editor assigned date:September 28,2022,PreQC No.I PIMP-22-14074 (PQ); Reviewed date: October 10, 2022, QC No. IPIMP-22-14074; Revised date:October 17, 2022, Manuscript No. IPIMP-22-14074 (R);Published date: October 25, 2022; DOI: 10.36648/2574-285x.7.5.21.
Citation: Emmanuel C (2022) Evaluation of Radiation Dose Received By Pediatrics Patients during Routine X –Ray Examinations of the Chest in Three Selected Hospitals in Cameroon. J Med Phys Appl Sci Vol. 7. No.5:21
The radiation risks associated with children are higher than the risks for adults. Children have growing organs and they have a longer life expectancy than that of adults. Consequently, the effects of damage from radiation could be greater in children than in adults. This study sought to measure the mean Entrance Surface Dose (ESD) and third quartile values for pediatrics chest X–ray examinations, compare these to findings and recommendations from other studies and propose methods of dose reduction in three hospitals (H1–H3) in Cameroon. The age groups considered in this study were <1 year, 1–< 5 years, 5–<10 years and 10–15 years. The mean ESD for the chest AP in the age range < 1year in the three hospitals (H1–H3) were respectively 0.10, 0.09, 0.66 mGy; for the age 1-<5years were respectively 0.12, 0.10, 0.48 mGy and for the age range 5-<10 years were respectively 0.13, 0.09, 0.34 mGy. According to NRPB 2000, the reference levels are given in defined age that is (newborn, 1 year, 5 years, and 10 years). For example, for AP and PA chest radiographs; 0.05 mGy for newborns and 1 year old, 0.07 mGy for 5 years and 0.12 mGy for 10 years old children. It is difficult to compare this reference with the results of this study where age range had been used to evaluate the mean ESD values but can still provide useful information for comparison. A weak correlation was also found between ESD and weight/age. The fluctuation of dose from one hospital to another is an indication that there is a need for optimization in pediatrics radiography. This can be done through Quality Assurance Programs (QAP).
Entrance surface dose; Pediatrics; Optimization
Living things have evolved in an environment that has significant levels of ionizing radiations. Radiation is used in medical diagnosis and treatment of diseases such as cancer. Thus, there are a lot of benefits from the careful use of radiations (X–rays). Despite these benefits, the use of medical X– rays still have hidden problems. The prevention of the hazardous effects of ionizing radiation has been a critical concern despite their valuable contribution in medical imaging to diagnosis and subsequent treatment of various diseases [1]. Radiation protection in pediatric radiology requires special attention than in adult radiology because children are more sensitive to radiation and at higher risk than adults [2] The higher risk in children is explained by their longer life expectancy, which allows more time for any harmful effects of radiation to manifest; and the fact that developing organs and tissues are more sensitive to the effects of radiation [3]. The importance of constant monitoring of the radiation dose received by pediatrics during a radiological examination cannot be understated as studies done indicate that risk for radiation related cancer in children is a result of early life exposure to ionizing radiation [4]. On this note, Quality Assurance Programs (QAP) should be implemented in Pediatric radiology. Quality assurance in pediatrics radiology is more important, since it is known that children are more sensitive to radiation than adults [5]. Therefore, in this case, closer attention should be paid to improve the diagnostic information, reducing the child dose as much as possible [6]. The use of Diagnostic Reference Levels (DRLs) is a very useful and applicable method in many countries as a measure to reduce over exposure or under exposure [7]. DRLs are set based on the survey of different hospitals using different X–ray tubes or machines. From the findings of the researchers’, there are no radiation dose survey measures, for both adults and pediatrics at the studied Hospitals and so therefore there is a possibility of over exposure. The National Agency for Radiation Protection (NARP) in Cameroon which is responsible for radiation protection of patients, general public and staff have not yet set baseline dose which are applicable to each hospital. A dosimetric survey was therefore important to ensure that children are not put at higher risk of developing radiation-induced cancer. The purpose of this article is to report on a study that sought to investigate the radiation dose received by children (1 to 15 years) in terms of ESD, evaluated using mathematical models from the data obtained from the patients and the exposure parameter from the X–ray machine during Exposure for pediatric chest, in three hospitals in Cameroon. From the results we shall establish a correlation between ESD, weight, height and age and comparing them with those reported in literature.
This study was carried out to estimate, using mathematical models, the ESD received by pediatric patients referred to diagnostic chest X–rays in three hospitals in Cameroon. These three hospitals were chosen because they receive a large number of patients. A total of 232 radiographs were obtained from children within the age of 1 week to 15 years, 115 females and 117 males. A total of three X–ray machines were used in this study with characteristics summarized in Table 1.
Hospital | Equipment model | Date of manufacture | Date of installation | Serial number | Film screen | Total filtration |
---|---|---|---|---|---|---|
Type/Speed | ||||||
H1 | GE Healthcare | May –2011 | 2012 | 75422HL6 | Fujifilm/400 | 2.5 mmAl |
H2 | Stepanix | May –2011 | 2012 | 11H411 | Fujifilm/400 | 3.09 mmAl |
H3 | CGR/ATAK25 | 1968 | 1974 | 6IR659 | Kodac/200 | 2.93 mmAl |
Table 1: Characteristics of the investigated X–rays machines.
The measurement of the tube output was performed using a calibrated DIAVOLT universal all in one Quality Control (QC) meter. It is a non-invasive X-ray meter for kV, dose and irradiation time measurements at X-ray installations. It was used in the absence of the patients in this study to obtain the tube output of the equipment. Two examination techniques were taken in account: PA and AP. Radiological parameters such as kilo Voltage (kV), product of X-ray tube current and time (mAs) and Focus-to-Skin Distance (FSD) with the patient characteristics such as sex, weight, height and age were recorded with the assistance of the technician or engineer on duty. The type of Xray examination and the view (projection) were obtained from the patient request form. The patient’s age was asked from patient care taker, the height was measured using a measuring tape and the weight of the patient was obtained using a balance. Information such as weight and age were also verified from their medical booklets. During the examination, the kV, and mAs were directly read from the control console and the tape which was fixed on the X-ray tube helped us to measure FSD directly after the patient was positioned for the examination. The ESD received by pediatrics patients was evaluated using equation 1[8] with a backscatter factor of 1.35 [9].
Where
OP is the output in mGy/mAs of the X–ray tube at 80 kV at distance of 100 cm normalized to 20 mAs
mAs is the time current product measured in mAs and
FSD is the focus to skin distance
The data collected was entered and analyzed using excels (2007). The results of the min, max, mean, 3rd quartile and standard deviation are presented in Table 2.
Weight/kg | Height/cm | kV | mAs | Age | |
---|---|---|---|---|---|
Minimum | 3.5 | 27 | 40 | 0.97 | 1day |
Mean | 15.75 | 163 | 75.51 | 6.59 | 2.56 |
Maximum | 62 | 74.86 | 120 | 30 | 15 |
Standard deviation | 11.11 | 32.85 | 26.85 | 7.18 | 3.64 |
Table 2: Summary of patient demographic and exposure parameters used in this study.
The male patients were 50.43% (n=117) of the total population while the female where 49.57% (n=115) of the total population. The patients were divided into four groups separate age groups: <1yr, 1-<5 yrs, 5-<10 yrs and 10–15 yrs. This was done due to the wide variation in body sizes common with pediatric patients. The separation of the patients into different age groups made it easy for the calculation of the ESD [10]. Patient demographic data and exposure parameters for this study are summarized in Table 2, while the results of the ESD obtained from this study for the chest PA and AP examinations are displayed in Table 3.
Projection | Studied centers | ESD (mGy) | |||||
---|---|---|---|---|---|---|---|
Mean | Min | Max | median | 3rdqtle | SD | ||
<1 years | |||||||
AP | H1 | 0.1 | 0.07 | 0.34 | 0.07 | 0.07 | 0.08 |
H2 | 0.09 | 0.03 | 0.2 | 0.09 | 0.1 | 0.04 | |
H3 | 0.66 | 0.35 | 1.06 | 0.67 | 0.75 | 0.17 | |
1–<5 years | |||||||
AP | H1 | 0.12 | 0.07 | 0.35 | 0.1 | 0.13 | 0.08 |
H2 | 0.1 | 0.06 | 0.14 | 0.11 | 0.12 | 0.04 | |
H3 | 0.48 | 0.19 | 0.86 | 0.49 | 0.57 | 0.23 | |
PA | H1 | 0.09 | 0.07 | 0.13 | 0.07 | 0.11 | 0.03 |
H2 | 0.1 | 0.06 | 0.17 | 0.09 | 0.12 | 0.04 | |
H3 | 0.32 | 0.18 | 0.49 | 0.32 | 0.35 | 0.08 | |
5–<10 years | |||||||
AP | H1 | 0.13 | 0.11 | 0.14 | 0.14 | 0.14 | 0.02 |
H2 | 0.09 | 0.06 | 0.16 | 0.08 | 0.08 | 0.04 | |
H3 | 0.34 | 0.3 | 0.38 | 0.32 | 0.36 | 0.03 | |
PA | H1 | 0.13 | 0.1 | 0.19 | 0.13 | 0.15 | 0.03 |
H2 | 0.08 | 0.03 | 0.14 | 0.07 | 0.08 | 0.04 | |
H3 | 0.35 | 0.19 | 0.48 | 0.36 | 0.37 | 0.06 | |
10-15 years | |||||||
PA | H1 | 0.2 | 0.14 | 0.26 | 0.19 | 0.22 | 0.03 |
H2 | 0.12 | 0.04 | 0.2 | 0.15 | 0.17 | 0.06 | |
H3 | 0.5 | 0.3 | 1.59 | 0.44 | 0.52 | 0.27 | |
AP | H2 | 0.13 | 0.11 | 0.14 | 0.14 | 0.14 | 0.02 |
Table 3: Statistical Distribution of ESD (in mGy) for the three hospitals and age ranges.
The ESD for chest AP was plotted against weight, age, mAs and kV and the scatter diagrams of figures 1 to 4 were obtained. From (Figures 1-4) it shows that there was no significant correlation between ESD and weight (r=-0.13292), between ESD and age (r=-0.2249) and between ESD and kV (r=-0.71256). However, 3 shows a significantly strong correlation between ESD and mAs (r=0.84564).
The ESD for of 232 patients were evaluated and the results presented in Table 3. A mean dose of 0.12 μGy was obtained for the chest AP for children in the age group <1 year. This value is slightly higher compared to a recommended value of 0.05 μGy by the NRPB (2000). The mean values of the ESD obtained from H1 and H2 for the age group 1-<5 years compare favorably with the recommended values by the NRPB and EC, while those from H3 were higher than the recommended values (see Table 4 and Figure 5) and those from similar literature [11].
Projection | This study | NRPB | EC | Nigeria, 2013 |
---|---|---|---|---|
-2000 | -1996 | (Ademola, A.K., et al. |
||
<1 year | ||||
Chest AP | 0.28 | 0.05 | ||
1–5 years | ||||
Chest AP | 0.23 | 0.07 | 0.1 | |
Chest PA | 0.17 | 0.07 | 0.1 | 0.11 |
5-< 10 years | ||||
Chest AP | 0.19 | 0.12 | ||
Chest PA | 0.19 | 0.12 | 0.159 | |
10–15 years | ||||
Chest AP | 0.13 | |||
Chest PA | 0.27 | 0.62 |
Table 4: Comparison of the mean ESD obtained from this study with reference.
The fluctuation in dose could be due to factors such as state of equipment and the technique or protocol used. The radiation dose received by patients is greatly influenced by the equipment used and the protocol applied. It can be seen from Table 1 that the X–ray machine of H3 is about 50 years old from the year of manufacture to the time of this study, meaning according to the suspension levels for diagnostic radiography, the tube of H3 should be suspended [12]. Besides, the film speed is extremely low; meaning high doses of radiations will be delivered to patients to have a good image quality. Thus, from all indications, patients will receive high doses than normal from H3. The quantity of radiation dose received by a patient is greatly affected by the radiographic technique used. Each hospital has its own protocol for every examination, for example at H3 the FSD was in the range 70–87 cm, meanwhile at H1 and H2 the FSD were respectively in the ranges 74–133 cm and 62–132 for the age group<1 year. This difference in FSD has an effect on patient dose and the doses reported by different authors. Low kV and high mAs technique are also attributing factors to dose received by a patient. A study carried out in Nigerian hospitals showed that those centers which used this technique obtained higher ESD values compared to those that used high kV and low mAs [13]. This is the other reason why ESD values were high at H3 as low kV (42–75 kV) and high mAs (5–30 mAs) values were used throughout the study. We expected from this study that as the weight/age of the patients increase, the ESD should also increase. This was not the case from our findings. There was a large fluctuation in ESD with weight/age (as seen in Figures 1 and 2). The correction coefficient between ESD and weight was (r=-0.13292) and between ESD and age it was (r=-0.2249). Thus there was no significant correlation between ESD and weight/ age. The fluctuation in dose weight/age could to due variation in patient size which is common with pediatric patients. Besides, there was no significant relationship between ESD and kV as the correlation coefficient between kV and ESD was found to be (r=-0.71256). However, there was a significantly strong correlation between mAs and ESD (r=0.84564), as expected.
Diagnostic radiology is the greatest source of public exposure to radiation. Although are medically justifiable, it is important to optimise the dose per examination [14] especially when dealing with pediatrics. Due to the cell sensitivity of children to ionizing radiation, special attention is required in radiation protection of Pediatrics patients than in adults. The United Nation Scientific Committee on Effects of Atomic Radiation (UNSCEAR) reported that children exposed to radiation at the early age of less than 5 years are twice to thrice sensitive to effects of radiation than adults [15]. These results suggest that pediatric patients especially below the age of 5 years may be receiving higher radiation dose compared to the recommended doses. This could be due to the lack of adequate quality control measures on the equipment and the age of the machines and radiographic techniques applied.
Results obtained in this study may act as a baseline for the establishment of local DRLs for chest examinations and they serve as a guideline in improving radiographic practice. However, due to the limitations in this study, the researchers recommend further studies in different hospitals as that may yield a more reliable mean ESD and third quartile value. The researchers also recommend careful implementation of the ALARA principle, application of more appropriate radiographic technique which is relevant to pediatrics patients, the setting up centres dedicated for pediatric examinations and to create awareness on suspension levels in diagnostic radiology.
[Crossref] [GoogleScholar][Indexed]
[Crossref] [GoogleScholar][Indexed]
[Crossref] [GoogleScholar][Indexed]
[Crossref] [GoogleScholar][Indexed]